NanoPhotometer
®
N120/NP80/N60/N50/C40 User Manual Version 4.3.1
74
11. Apply the blank ddH
2
0 or buffer to the illuminated sample window on
pedestal for the reference measurement and select blank to initiate the
reading.
Note: The illumination of the sample window can be switched off in the
preferences.
12. Use a lint-free laboratory wipe to clean both the sample window on
pedestal and mirror in the lid arm prior to applying the next sample.
Note: It can be helpful to apply the blank a second time and read it as a
sample to ensure a proper blank.
13. Apply sample to the sample window on pedestal and press the sample
button to initiate the measurement.
If auto sample is active the sample measurement is automatically started
after closing the lid arm.
Note: Auto sample function is only available NanoVolume methods and
new NanoPhotometer
®
versions it may not be available after updating an
older firmware version.
MEASUREMENT PROTOCOL MULTI SAMPLE
1. Select the Protein UV icon on the home screen.
2. To change between single and multi-sample application use the toggle
switch Single Sample in the parameter area.
3. To change the protein type push on BSA and a list with available options opens on the
right side.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Implen NanoPhotometer NP80 and is the answer not in the manual?

Implen NanoPhotometer NP80 Specifications

General IconGeneral
BrandImplen
ModelNanoPhotometer NP80
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals