EasyManua.ls Logo

Implen NanoPhotometer NP80 - Page 80

Implen NanoPhotometer NP80
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NanoPhotometer
®
N120/NP80/N60/N50/C40 User Manual Version 4.3.1
80
C =A
280
* ε
prot
* Ð
C Concentration in mg/ml
A
280
Absorbance at 280 nm (10 mm path)
A
BKG
Absorbance at selected background wavelength (10 mm path)
A
max,dye
Absorbance value at the absorbance maximum of the dye (10 mm path)
ε
prot
Extinction coefficient/protein factor in g*cm/l
cf
dye
Dye-dependent correction factor at 280 nm
Ð Dilution factor
Dye Concentration
For dye-labeled proteins, the concentration of the dye is calculated using a modified form of
the Beer-Lambert equation. For these calculations, the instrument considers the absorption
maximum of the dye, and a dye-specific extinction coefficient (see Table 4 on page 82). The
dye concentration is calculated with or without background correction for are as follows:
With background correction:
C =
󰇡(A
max,dye
A
BKG
) * Ð󰇢
ε
dye
* 10
6
Without background correction:
C =
A
max,dye
* Ð
ε
dye
* 10
6
C Concentration
A
max, dye
Absorbance at the max dye absorbance value (10 mm path)
A
BKG
Absorbance at selected background wavelength (10 mm path)
ε
dye
Extinction coefficient of dye in M
-1
*cm
-1
Ð Dilution factor
Degree of Labeling (DOL)
DOL is the degree of labeling based on the average number of dye molecules coupled to a
protein molecule. The degree of labeling can be determined from the absorption spectrum of
the labeled antibody with or without background/dye correction as follows:
With background and with dye correction:
DOL =
(A
max, dye
A
BKG
) * ε
prot
󰇡A
280
A
BKG
cf
dye
* (A
max, dye
A
BKG
)󰇢 ε
dye
With background and w/o dye correction:
DOL =
(A
max, dye
A
BKG
* ε
prot
(A
280
A
BKG
) * ε
dye

Table of Contents