4Description 15
Translation of the original operating instructions Ecotec E3000, kina22en1-19, 1507
Response time
3 m sniffer line 0.7 sec.
5 m sniffer line 0.9 sec.
10 m sniffer line 1.4 sec.
15 m sniffer line 3.0 sec.
Table 3: Factory settings
Alarm profile Trigger alarm
Alarm delay Deactivated
Number of measuring points (I•Guide) 4
Functions Activated
Internal Activated
Recording output Auto
Baud rate and blank flange 9600 CR+LF
Pressure unit mbar
Flow rate
Lower limit 100 sccm
Upper limit 250 sccm
Sensitivity test Activated
Gas, definition Gas 1, gas 2, gas 3, gas 4, gas 5, gas 6
Device speaker Activated
Handle speaker Trigger value
Selecting a cathode A
I•Guide Deactivated
Calibration (cal), internal Activated
Contrast Not inverted, setting 30
Volume 2
Minimum volume 2
Leak rate, selected, highest Automatic
Leak rate filter I•Filter
Menu PIN Deactivated, 0000
Measuring time (I•Guide) 1 second
Measuring mass 69
Peak value Deactivated, 5 seconds
ECO-Check Activated
Relay outputs See interface description
RS-232 protocol ASCII
Sniffer probe, light Activated, Level 4
Sniffer probe, filter, maintenance 100 hours
Recorder, gas Auto
Scaling of the recorder Logarithmic
Language English
SPS outputs and inputs See interface description
Table 2: Technical data (cont.)