34 6 Operation
Translation of the original operating instructions Ecotec E3000, kina22en1-19, 1507
RS-232 protocol ASCII
Diagnosis
Printer Auto
Printer Manual
ECO-Check If you do not use an ECO-Check calibrated leak, you should choose “Deactivate” here.
Otherwise, warning 71 “No communication with ECO-Check” will be issued at every start
of the Ecotec E3000.
If you use an ECO-Check calibrated leak, you can set up a warning for the intended time of
use here: 14/30/60/90 days.
6.4 Settings for the measurements
With delivery, the data are programmed for the following gases and the measurement
results are shown below each other in the display:
R134a
R22
R600a
He
The gas selection can be changed at any time.
6.4.1 Selecting the gas, changing gas parameters, activating
measurement
You can access the settings via the main menu
Select “Measuring parameters”
Fig. 19: Gas information for the measurement
The display shows:
four gases
the respective mass position
the respective trigger value
the addition “Deactivated” if the associated gas is currently not searched for, see below.
You can also define your own gas with the key “Gas def., see “6.4.5 Setting a user-defined
gas”, page 42.
Press the key to the right of the gas whose parameters you want to change. The window
“Measuring param. gas …” will open.

Table of Contents

Other manuals for Inficon Ecotec E3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon Ecotec E3000 and is the answer not in the manual?

Inficon Ecotec E3000 Specifications

General IconGeneral
BrandInficon
ModelEcotec E3000
CategorySecurity Sensors
LanguageEnglish

Related product manuals