User’s manual of CLEO Patient Monitor
- 24 -
ALARM SWITCH
ON and OFF for choice, the factory setting is ON.
If the NIBP value is above or below the NIBP alarm limit, and the settings are set to ON,
then the alarm will be activated. When the settings are set to OFF, the alarm indicator will
not light up; the corresponding alarm parameters will not flash.
PRESSURE UNIT
mmHg or kPa, the factory setting is mmHg.
PATIENT TYPE
ADULT TYPE
This setting applies to the adult mode. Inflate the cuff to 180mmHg (24kPa). If the NIBP
value cannot be measured, then inflate the cuff an additional 50mmHg (6.7kPa), but note
that the maximum pressure cannot exceed 280mmHg (37.3kPa), and the enduring
pressure range is 50-280mmHg. The factory setting use ADULT TYPE.
PEDIATRIC/NEONATE TYPE
This setting applies to the PEDDIATRIC or NEONATE mode. Inflate the cuff to 60mmHg
(8kPa). If the NIBP value cannot be measured, then inflate the cuff an additional 30mmHg
(4kPa), But note that the maximum value cannot exceed 150mmHg (20kPa), and the
enduring pressure range is 50-150mmHg.
If this setup is done before the NIBP module is initiated, the settings will not be effective.
The inflating ranges shown above have been implemented on NIBP, which uses them to
guarantee the safety of the patient.
INFLATION MODE
There are three choices: Manual, Auto and STAT.
MANUAL MODE
Press the START/STOP button to begin inflation. The information area of the display
should read “Manual measuring…”, which shows that the measurement is being taken.
Once the NIBP measurement is finished, NIBP parameter area will display the values and
the information area will read “Manual measuring ended!, and the measurement process
will have finished.
If the NIBP value cannot be measured, the NIBP parameter area will display any error
messages and automatically attempt to take the measurement again up to three times. If
the value still cannot be measured, the information area will readRETRY OVER, and no
more measurements will be taken.
During the measurement, press the START/STOP button again to stop the NIBP
measurement process. The information area will read “Stop Manual measuring.”
AUTOMATICAL MODE
NIBP parameter area will display the countdown of "Auto measuring..." (TIME INTERVAL).
Under this setting, the machine will continue re-measure NIBP at every time interval until
the mode is changed.
If you start a measurement manually, the monitor will then continue automatically repeat
NIBP measurements at the set time interval.
If the NIBP measurement is finished, the NIBP parameter area will display the values and