- 25 -
the information indicating area will read "Auto measuring ended
!
". The monitor will then
continue automatically measuring until the mode is changed.
If the NIBP value cannot be measured, the NIBP parameter area will display any error
messages and will attempt to take the measurement again up to three times. If the value
still cannot be measured, the information area will read “RETRY OVER”. And
automatically continue to take measurements at every time interval until the mode is
changed.
If the START/STOP button is pressed during any point in the countdown period, the
monitor will immediately begin a new measurement.
During the measurement, pressing the START/STOP button again will stop the current
NIBP measurement, and the information area will read "Stop Auto measuring", but the
monitor will continue to automatically take measurements at every time interval.
STAT MODE
In the stat mode, the monitor will measure NIBP three times without stopping and then will
stop. You can also press the START/STOP button to end the measurement manually.
Press the START/STOP button to begin inflation. The information area will display "STAT
measuring..." to indicate that the measurement is being taken. Once the NIBP
measurement has finished, the NIBP parameter area will display the values and the
information area will read "STAT measuring ended".
If the NIBP value cannot be measured, the NIBP parameter area will display any error
messages and automatically will attempt to take the measurement again up to three times.
If the value still cannot be measured, the information area will read “RETRY OVER
!
”, and
no more measurements will be taken.
During the measurement, if you press the START/STOP button again, the information
area will read "Stop STAT measuring", and the monitor will stop the NIBP measurement.
TIME INTERVAL
This setting is used by the automatic inflation mode. You can input the time interval
you want, as long as it is between 1 min to 4 hours.
The measured value will be display on the NIBP parameter area for 240 minutes
unless a new measurement is taken during this period. On the appropriate trend graph
and trend table, the parameter will be included for the appropriate length of time.
Prolonged non-invasive blood pressure measurements in Auto mode may be
associated with purport, isocheimal and neuropathy in the limb wearing the cuff. When
monitoring a patient, examine the extremities of the limb frequently for normal color,
warmth and sensitivity. If any abnormality is observed, stop the blood pressure