EasyManuals Logo

Instron 34SC-1 User Manual

Default Icon
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
Chapter: Maintenance
112 M10-17313-EN
Daily Maintenance Checks
Before operating the system each day, ensure:
All cable connections are tight and secure.
All grips, fixtures and accessories are free of dirt, damage and deformation.
The load frame is level. If an adjustment is necessary, refer to “Level the load
frame” on page 31.
Signal and power cables have adequate slack to prevent excessive strain on
connectors.
All cables and pneumatic hoses are free of wear and chafing. Re-route them if
necessary, and replace any that are damaged.
After turning on the system, make sure that power is adequately supplied to the
electronics.
Correct any problems before you operate the testing system. If you require assistance,
contact Instron
®
Service.
Periodic Inspections
Every six to twelve months, perform the following inspections:
Visually inspect the machine for any loose fittings. Check the limit stops, cable
connections, and connections for any accessories that are attached to the machine.
Tighten any loose connections that you may find.
Operate the crosshead through the full range of motion. It should move smoothly
with no unusual noise, or erratic motion. If it does not, refer to Table 10 on
page 114 for lubrication recommendations.
Test the limit stops to ensure they are in working order. Refer to “Test Limit Stops”
on page 118.
If you have any additional safety equipment added to the machine, test the
equipment to ensure it is in working order.
If you notice any problems resulting from these inspections, contact Instron
®
Service for
immediate assistance.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Instron 34SC-1 and is the answer not in the manual?

Instron 34SC-1 Specifications

General IconGeneral
BrandInstron
Model34SC-1
CategoryTest Equipment
LanguageEnglish

Related product manuals