EasyManuals Logo

Instron 34SC-1 User Manual

Default Icon
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
27
Residual risks
Product Support: www.instron.com
Use a specimen insertion tool and other appropriate equipment and practices to
keep fingers out of the space between the fixtures.
In addition, if the gap between grips and fixtures is less than 25 mm:
Use a specimen insertion tool to keep fingers out of the space between the fixtures.
Use an interlocked shield to limit or disallow motion when the shield door is open.
When the space is less than 25mm, the crush hazard is very significant, so this is
the best option.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Instron 34SC-1 and is the answer not in the manual?

Instron 34SC-1 Specifications

General IconGeneral
BrandInstron
Model34SC-1
CategoryTest Equipment
LanguageEnglish

Related product manuals