Chapter: Installation
48 M10-17313-EN
2. Turn the power switch on the machine to the On (1) position.
The white LED above the
DISABLED indicator flashes.
3. Turn on the power to the Bluehill
®
Operator Dashboard and any other accessories
that make up the testing system.
4. Start the Bluehill
®
software.
The first time the software is started, the configuration wizard runs.
5. Follow the instructions in the configuration wizard. The keycode for the software is
printed on the media supplied. Steps include:
• select the controller type and frame model
• enter the system ID (located on a label on the rear of the frame)
• enter the Ethernet address (located on a label on the controller panel)
• enter any additional information, e.g. service contract number
• select any additional accessories that are installed
If you need to change any of these settings, when you purchase an additional accessory
for example, go to the Admin tab in the software.