ASSIST PLUS – Operating instructions V02 Programming
28 INTEGRA Biosciences AG
Press to save your settings. This will return you to the list of Serial Dilution programs.
Advan-
ced
Tip Travel Defines the distance which the tips move during an aspira-
tion, dispense and mix step to artificially track the liquid level,
see “7.8
Tip travel distances” on page 47.
• Source Asp: moves the set distance down during aspira-
tion in the source location.
• Source Mix: moves the set distance down and up during
mixing in the source location.
• Target Disp: moves the set distance up during dispensing
in the target location. The same distance is used to move
down during aspiration in the target location.
• Target Mix: moves the set distance down and up during
mixing in the target location.
Options Steps Description of Serial Dilution