SERVICE INSTRUCTIONS
25
Remote status bar
indicator LED blink
code
Servo light
module LED
blink code
Cause Fault remedy
11
Maximum current time-out
(servo motor overloaded or
overheated)
Switch the electronics
system off and on again.
12
-
Compatibility problems
between ACS modules
Remove all electronic
modules except the
Power Module and the
Remote. Re-attach
modules one by one to
determine which one is
causing the fault. Remove
incorrect module See
"
Replacing electronics
components
" on page 77