SERVICE INSTRUCTIONS
Invacare
®
- G50
24
Remote status bar
indicator LED blink
code
Servo light
module LED
blink code
Cause Fault remedy
6 -
Magnetic brake defective or
drive disengaged
Engage motor. Switch the
electronics system off and
on again.
Check the connecting
plug
Replace motor See
"
Replacing a motor" on
page
41
7 -
Battery voltage too low
Charge battery. See User'
Manual
7
Battery voltage too low (below
17V)
Check battery
connections
Check fuses.
Check batteries and
replace if necessary. See
User' Manual
8 -
Battery voltage too high
Turn on lighting
9
CANL error
Check bus cable/data
cable for short circuit. An
open circuit or a short
circuit in one of the ACS-
Modules can also cause
this error. See "
Checking
the cables
" on page 87
and "
Replacing
electronics
components
" on page 77
9
CANL error problems on the
CANL channel /
communication over CANL not
possible.
Check bus cable for
continuity
10 -
CANH error
Check bus cable/data
cable for short circuit. An
open circuit or a short
circuit in one of the ACS-
Modules can also cause
this error. See "
Checking
the cables
" on page 87
and "
Replacing
electronics
components
" on page 77
10
CANH error problems on the
CANH channel /
communication over CANH
not possible.
Check bus cable for
continuity
11 -
Maximum current time-out
(drive overloaded or
overheated)
Switch the electronics
system off and on again.