TempScan / MultiScan User's Manual Chapter 16: API Command Reference 327
S
- Set Real-Time Clock
TYPE
System
EXECUTION
Immediate
SYNTAX
Stime
Set the internal real-time clock where
time
is in the standard Time/Date
input format:
hh:mm:ss:t,MM/DD/YY
where
hh
is in hours,
mm
is in
minutes,
ss
is in seconds,
t
is in tenths of a second,
MM
is the month,
DD
is the day, and
YY
is the year.
S?
Query the internal real-time clock.
DESCRIPTION
The Set Real-Time Clock (
S
) command is used to set the battery-backed-up internal real-time clock of the unit.
The
time
argument follows the same format as the standard Time/Date input format. In order for the Time/Date
stamping of output data to be accurate, the real-time clock must be set to an accurate Time/Date. Once set, the
real-time clock will remember the current time, even when powered off.
For more information on the real-time clock, Time/Date stamping and data formats, see section Additional
Configuration, section Stamping Configuration and section Data Format Configuration, respectively, in the
chapter System Configuration.
EXAMPLE
PRINT#1, “OUTPUT07; S14:00:00.0,4/30/97X”