Technical Specification
Parameter IT6018C-300-225
Maximum Input Ap-
parent Power
19.8kVA
Frequency
47Hz~63Hz
Setup Stability-30min Voltage
≤0.02%+0.02%FS
±(%of Output+Offset)
Current ≤0.1% + 0.1%FS
Setup stability-8h Voltage
≤0.02%+0.02%FS
±(%of Output+Offset)
Current ≤0.1% + 0.1%FS
Readback Stability-30min Voltage
≤0.02%+0.02%FS
±(%of Output+Offset)
Current ≤0.1% + 0.1%FS
Readback stability-8h Voltage
≤0.02%+0.02%FS
±(%of Output+Offset)
Current ≤0.1% + 0.1%FS
Efficiency
~92%
Remote Sense Compensa-
tion Voltage
≤5V
Command Response Time
2mS
Power Factor 0.99
Storage Temperature
-10℃~70℃
Protective Function
OVP, OCP, OPP, UVP, UCP, OTP, Vsense protection
Standard Interface
Standard: USB, CAN, LAN, VCP; optional: GPIB, analog card
(includes RS232), fiber optic socket
Isolation ( output to ground)
1000V
Working Temperature
0~50℃
Dimension ( mm) 483mm(W)*801.61mm(D)*151.3mm(H)
Weight( net)
40KG
Note:
1. The instrument needs to be customized under other AC input voltage ranges. For models
equipped with an emergency stop module (including AC contactor), the AC input voltage range
can only be: 342V~528V
Load Mode:
Parameter IT6018B-300-225
Rated value ( 0
℃-50 ℃)
Input voltage
0~300V
Input current
0~225A
Input power
0~18kW
Copyright © Itech Electronic Co., Ltd.
250