Technical Specification
Parameter IT6018B-800-75
±(%of Output+Offset)
Current ≤0.1% + 0.1%FS
Setup stability-8h Voltage
≤0.02% + 0.02%FS
±(%of Output+Offset)
Current ≤0.1% + 0.1%FS
Readback Stability-30min Voltage
≤0.02% + 0.02%FS
±(%of Output+Offset)
Current ≤0.1% + 0.1%FS
Readback stability-8h Voltage
≤0.02% + 0.02%FS
±(%of Output+Offset)
Current ≤0.1% + 0.1%FS
Efficiency
~92%
Remote Sense Compensa-
tion Voltage
≤8V
Command Response Time
2mS
Power Factor 0.99
Storage Temperature
-10℃~70℃
Protective Function
OVP, OCP, OPP, UVP, UCP, OTP, Vsense protection
Standard Interface
Standard: USB, CAN, LAN, VCP; optional: GPIB, analog card (in-
cludes RS232), fiber optic socket
Isolation ( output to ground)
1500V
Working Temperature
0~50℃
Dimension ( mm) 483mm(W)*801.61mm(D)*151.3mm(H)
Weight( net) 40kg
Note: 1. The instrument needs to be customized under other AC input voltage ranges. For mod-
els equipped with an emergency stop module (including AC contactor), the AC input voltage
range can only be: 342V~528V
Load Mode:
Parameter IT6018B-800-75
Rated value ( 0
℃-50 ℃)
Input voltage
0~800V
Input current
0~75A
Input power
0~18kW
Input resistance
0.001~7500Ω
Min. operating
voltage
2.1V at 75A
Setup Resolution Voltage
0.01V