Technical Specification
Parameter IT6018B-2250-25
Isolation ( output to ground)
3000V
Working Temperature
050
Dimension ( mm) 483mm(W)*801.61mm(D)*151.3mm(H)
Weight( net)
40KG
Note: 1. The instrument needs to be customized under other AC input voltage ranges. For mod-
els equipped with an emergency stop module (including AC contactor), the AC input voltage
range can only be: 342V528V
Load Mode:
Parameter IT6018B-2250-25
Rated value ( 0
-50 )
Input voltage
02250V
Input current
025A
Input power
018kW
Input resistance
0.0017500Ω
Min. operating
voltage
6.25V at 25A
Setup Resolution
Voltage
0.1V
Current 0.001A
Power 0.001kW
Resistance 0.001Ω
Read Back
Resolution
Voltage
0.1V
Current 0.001A
Power 0.001kW
Resistance 0.001Ω
Setup Accuracy
(within 12 months,
25 ±5) ±(
ofOutput+Offset)
Voltage
≤0.02% + 0.02%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Resistance ≤1% + 1%FS
Read Back Accu-
racy (within 12
months, 25 ±5)
±(of Output
+Offset)
Voltage
≤0.02% + 0.02%FS
Current ≤0.1% + 0.1%FS
Power ≤0.5% + 0.5%FS
Resistance ≤1% + 1%FS
Setup Temperature
Coefficient ±(of
Output/+Offset)
Voltage
≤50PPM/
Current ≤200PPM/
Copyright © Itech Electronic Co., Ltd.