Power Supply Function
OCP
Overcurrent protection
Off
Turn the OCP function off. (Def)
On Turn the OCP function on.
Level OCP limit
Delay
Protection delay time,
see Protection Delay.
OPP
Overpower protection
Off
Turn the OPP function off. (Def)
On Turn the OPP function on.
Level OPP limit
Delay
Protection delay time,
see Protection Delay.
UCP
Undercurrent protection
Off
Turn the UCP function off. (Def)
On Turn the UCP function on.
Warm-up
Indicates the instrument
warm-up time. This time
is set to prevent the in-
strument from triggering
protection when the cur-
rent is rising. Because
this transient condition
should not be consid-
ered as an undercurrent
fault and there is no
need to trigger
protection.
Level UCP limit
Delay
Protection delay time,
see Protection Delay.
UVP
Undervoltage protection
Off
Turn the UVP function off. (Def)
On Turn the UVP function on.