Input Settings 
  Copyright © ITECH Electronic Co., Ltd.  28 
Examples         
CURR:SLEW FALL 3 
Query Syntax       
[SOURce:]CURRent:SLEW:FALL? 
Returned Parameters     
<NR2> 
[SOURce:]CURRent:SLEW:FALL? 
This command is used to query the slew rate of the current for negative going 
transitions. 
Command Syntax 
[SOURce:]CURRent:SLEW:FALL? 
Parameters 
None 
Unit 
A/uS 
*RST Value 
MAXimum 
[SOURce:]CURRent:PROTection[:LEVel] <NRf+> 
This command sets the current protection level. If the input current exceeds the 
current protection level for the time specified by CURR:PROT:DEL, the input is 
turned off. 
 
NOTE: Use CURR:PROT:DEL to prevent momentary current limit conditions 
caused by programmed changes from tripping the overcurrent protection. 
 
Command Syntax     
[SOURce:]CURRent:PROTection[:LEVel] <NRf+> 
Parameters       
MIN ~ MAX | MINimum | MAXimum 
Unit           
A 
*RST Value       
MAXimum