Chapter10 Timing Test Commands
TIMing[:STATe]
This command is used to start or stop the Timing test.
After sending the TIM 1 command, the front panel of the instrument begins to
display the timing of the test. Unit: S;
After sending the TIM 0 command, the front panel timing function disappears.
Command Syntax
TIMing[:STATe] <bool>
Parameters
0|1|OFF|ON
Examples
TIM ON
Query Syntax
TIMing[:STATe]?
Returned Parameters
0|1
TIMing:LOAD:SETTing
This command is used to determine whether to change the load setting during
the Timing test.
Command Syntax
TIMing:LOAD:SETTing <bool>
Parameters
0|1|OFF|ON
When it is set to ON, after the test is started, the load is changed according to
the setting of TIMing:LOAD, and after the test is ended, the load input is turned
off; if it is set to OFF, the test is started/stopped without changing the load
setting.
Examples
TIM:LOAD:SETT OFF
Query Syntax
TIMing:LOAD:MODE?

Table of Contents

Other manuals for ITech IT8500plus Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8500plus Series and is the answer not in the manual?

ITech IT8500plus Series Specifications

General IconGeneral
BrandITech
ModelIT8500plus Series
CategoryTest Equipment
LanguageEnglish

Related product manuals