Timing Test Commands
Copyright © ITECH Electronic Co., Ltd. 61
Examples
TIM:TST:LEV 1
Query Syntax
TIMing:TSTart:LEVel?
Returned Parameters
<NR2>
TIMing:TEND:SOURce
This command sets the trigger source for ending the test.
Command Syntax
TIMing:TEND:SOURce <source>
Parameters
VOLT|CURR|EXT
Examples
TIM:TEND:SOUR VOLT
Query Syntax
TIMing:TEND:SOURce?
Returned Parameters
<source>
TIMing:TEND:EDGE
This command is used to set the trigger edge of the end test.
Command Syntax
TIMing:TEND:EDGE <edge>
Parameters
RISE|FALL
Examples
TIM:TEND:EDGE RISE
Query Syntax
TIMing:TEND:EDGE?
Returned Parameters
<edge>