Input Settings 
  Copyright © ITECH Electronic Co., Ltd.  29 
Examples         
CURR:PROT 3 
Query Syntax       
[SOURce:]CURRent:PROTection[:LEVel]? 
Returned Parameters     
NR2 
[SOURce:]CURRent:PROTection[:LEVel]? 
This command is used to query the current protection level.   
Command Syntax 
[SOURce:]CURRent:PROTection[:LEVel]? 
Unit 
A 
*RST Value 
MAXimum 
[SOURce:]POWer:PROTection[:LEVel] <NRf+> 
This command sets the power protection level. If the input power exceeds the 
power protection level for the time specified by POW:PROT:DEL, the input is 
turned off. 
 
NOTE: Use POW:PROT:DEL to prevent momentary power limit conditions 
caused by programmed changes from tripping the over power protection. 
 
Command Syntax 
[SOURce:]POWer:PROTection[:LEVel] <NRf+> 
Parameters 
MIN ~ MAX | MINimum | MAXimum 
Unit 
W 
*RST Value 
MAXimum(high range) 
Examples 
POW:PROT 100