Timing Test Commands
Copyright © ITECH Electronic Co., Ltd. 59
Returned Parameters
<mode>
TIMing:LOAD:MODE
This command is used to set the load mode during Timing test.
Command Syntax
TIMing:LOAD:MODE <mode>
Parameters
CURR|VOLT|POW|RES
Examples
TIM:LOAD:MODE CURR
Query Syntax
TIMing:LOAD:MODE?
Returned Parameters
<mode>
Related Commands
If TIM:LOAD:SETT OFF is executed, the setting of this command does not
take effect.
TIMing:LOAD:VALue
This command is used to set the load parameters for the Timing test.
Command Syntax
TIMing:LOAD:VALue <Nrf+>
Parameters
A / V / W / Ω, depending on TIMing:LOAD:MODE
Examples
TIM:LOAD:VAL 1
Query Syntax
TIMing:LOAD:VALue?
Returned Parameters
<NR2>
Related Commands
If TIM:LOAD:SETT OFF is executed, the setting of this command does not
take effect.