2-12 Return to Section Topics 2100-900-01 Rev. D / September 2011
Section 2: Getting Started Model 2100 6 1/2-Digit Resolution Digital Multimeter User’s Manual
Figure 2-17
Replacing the fuse
Step 6: Reinsert the voltage selector
Insert the voltage setting selector back into the socket and close the cap as shown in Figure 2-18.
CAUTION Verify that the correct voltage setting appears in the red voltage selection
window in the power module before powering up the unit. Refer to
Connecting the power” earlier in this section for instructions on how to
correctly power up the unit.
Figure 2-18
Reinserting the voltage selector