EasyManuals Logo

Keithley 3706A Reference Manual

Keithley 3706A
936 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #235 background imageLoading...
Page #235 background image
Section 6: Reading buffers Series 3700A System Switch/Multimeter Reference Manual
6-8 3700AS-901-01 Rev. D/June 2018
Reading buffer designations
To access the buffer, include the buffer attribute in the respective command. For example, the
following commands store five readings from the DMM into a buffer named readingbuffer:
-- Sets how many readings to take with the dmm.measure command.
dmm.measurecount = 5
-- Takes the measurements and stores them in readingbuffer.
dmm.measure(readingbuffer)
Do not use quotation marks around the reading buffer name when you send the
dmm.measure(readingbuffer) command from the instrument front panel, because a data type
error message will be logged to the error queue.
Buffer storage control attributes
Buffer storage attributes are summarized in the following table. To control which elements are stored
in the buffer, enable the attribute for the buffer (which sets it to 1). The following attributes are all
available per reading buffer. For example, to access the appendmode attribute for a buffer named
rb, send rb.appendmode.
Attribute
Description
appendmode
When off, a new measurement to this buffer will clear the previous contents
before storing the new measurement. When on, the first new measurement will
be stored at what was formerly rb[n+1].
This attribute is initialized to off when the buffer is created over the bus.
However, the default is on for the front panel or web interface to allow triggered
readings to fill a buffer without clearing the previous ones.
cachemode
When this attribute is on, the reading buffer cache improves access speed to
reading buffer data. When running successive operations that overwrite reading
buffer data without running any commands that automatically invalidate the
cache, the reading buffer may return stale cache data. This attribute is
initialized to on when the buffer is created.
collectchannels
When on, channel or channel pattern information is stored with readings in the
buffer. This requires eight extra bytes of storage per reading.
This value, off or on, can only be changed when the buffer is empty (cleared).
When the buffer is created, this attribute is initialized to on.
collecttimestamps
When on, timestamps will be stored with readings in the buffer. This requires
eight extra bytes of storage per reading.
This value, off or on, can only be changed when the buffer is empty (cleared).
When the buffer is created, this attribute is initialized to on.

Table of Contents

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
Model3706A
CategorySystem Switch/Multimeter
Voltage Range100 nV to 300 V
Basic DC Voltage Accuracy0.0035%
Basic Resistance Accuracy0.01%
Maximum Channels576
Maximum Voltage300 V
ConnectivityLAN, USB, GPIB
InterfaceLAN, USB, GPIB
Switching MatrixYes
Channels576 (6-slot mainframe)
Maximum Current3 A

Related product manuals