Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-264 3700AS-901-01 Rev. D/June 2018
lan.config.subnetmask
This attribute contains the LAN subnet mask.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
LAN restore defaults
Nonvolatile memory
"255.255.255.0"
Usage
mask = lan.config.subnetmask
lan.config.subnetmask = "mask"
mask
String that specifies the LAN subnet mask value in dotted decimal notation
Details
This attribute specifies the LAN subnet mask that will be used when the manual configuration method
is used to configure the LAN. This setting is ignored when DLLA or DHCP is used.
This attribute does not indicate the actual setting presently in effect. Use the
lan.status.subnetmask attribute to determine the present operating state of the LAN.
Example
print(lan.config.subnetmask)
Outputs the LAN subnet mask, such as:
255.255.255.0
Also see
lan.restoredefaults() (on page 11-265)
lan.status.subnetmask (on page 11-272)
lan.lxidomain
This attribute contains the LXI domain.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
LAN restore defaults
Nonvolatile memory
0
Usage
domain = lan.lxidomain
lan.lxidomain = domain
domain
The LXI domain number (0 to 255)
Details
This attribute sets the LXI domain number.
All outgoing LXI packets are generated with this domain number. All inbound LXI packets are ignored
unless they have this domain number.
Example
print(lan.lxidomain)
Displays the LXI domain.
Also see
None

Table of Contents

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
Model3706A
CategorySystem Switch/Multimeter
Voltage Range100 nV to 300 V
Basic DC Voltage Accuracy0.0035%
Basic Resistance Accuracy0.01%
Maximum Channels576
Maximum Voltage300 V
ConnectivityLAN, USB, GPIB
InterfaceLAN, USB, GPIB
Switching MatrixYes
Channels576 (6-slot mainframe)
Maximum Current3 A

Related product manuals