EasyManuals Logo

Keithley 3706A Reference Manual

Keithley 3706A
936 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #717 background imageLoading...
Page #717 background image
Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-338 3700AS-901-01 Rev. D/June 2018
schedule.alarm[N].enable
This attribute enables or disables an alarm.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Reset
Recall setup
Not saved
0 (schedule.OFF)
Usage
state = schedule.alarm[N].enable
schedule.alarm[N].enable = state
state
Disable the alarm(schedule.OFF or 0)
Enable the alarm (schedule.ON or 1)
N
Alarm number (1 or 2)
Details
If you enable an alarm that has a start time that is in the past, the alarm executes immediately.
If an alarm time in the past is used to start a scan, the alarm time may be missed by the scan start.
This can occur because the scan clears any pending triggers before it begins, so it will miss any
trigger generated from the alarm enable. To prevent a missed alarm, start the scan in the
background, then enable the alarm.
Example
schedule.alarm[1].enable = 1
Enables alarm 1.
Also see
schedule.alarm[N].EVENT_ID (on page 11-338)
schedule.alarm[N].fractionalseconds (on page 11-339)
schedule.alarm[N].period (on page 11-340)
schedule.alarm[N].ptpseconds (on page 11-340)
schedule.alarm[N].repetition (on page 11-341)
schedule.alarm[N].seconds (on page 11-341)
schedule.disable() (on page 11-342)
schedule.alarm[N].EVENT_ID
This constant describes the trigger event generated by the alarm N.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID = schedule.alarm[N].EVENT_ID
eventID
The trigger event number
N
Alarm number (1 or 2)

Table of Contents

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
Model3706A
CategorySystem Switch/Multimeter
Voltage Range100 nV to 300 V
Basic DC Voltage Accuracy0.0035%
Basic Resistance Accuracy0.01%
Maximum Channels576
Maximum Voltage300 V
ConnectivityLAN, USB, GPIB
InterfaceLAN, USB, GPIB
Switching MatrixYes
Channels576 (6-slot mainframe)
Maximum Current3 A

Related product manuals