EasyManuals Logo

Keithley 3706A Reference Manual

Keithley 3706A
936 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #808 background imageLoading...
Page #808 background image
Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-429
Details
Releases a trigger that was asserted with an indefinite pulse width. It also releases a trigger that was
latched in response to receiving a synchronous mode trigger.
Example
tsplink.trigger[3].release()
Releases trigger line 3.
Also see
tsplink.trigger[N].assert() (on page 11-424)
tsplink.trigger[N].clear() (on page 11-425)
tsplink.trigger[N].mode (on page 11-426)
tsplink.trigger[N].overrun (on page 11-427)
tsplink.trigger[N].pulsewidth (on page 11-428)
tsplink.trigger[N].stimulus (on page 11-430)
tsplink.trigger[N].wait() (on page 11-431)
tsplink.trigger[N].reset()
This function resets some of the TSP-Link trigger attributes to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.trigger[N].reset()
N
The trigger line (1 to 3)
Details
The tsplink.trigger[N].reset() function resets the following attributes to their factory
defaults:
â–ª tsplink.trigger[N].mode
â–ª tsplink.trigger[N].stimulus
â–ª tsplink.trigger[N].pulsewidth
This also clears tsplink.trigger[N].overrun.
Example
tsplink.trigger[3].reset()
Resets TSP-Link trigger line 3
attributes back to factory default
values.
Also see
tsplink.trigger[N].mode (on page 11-426)
tsplink.trigger[N].overrun (on page 11-427)
tsplink.trigger[N].pulsewidth (on page 11-428)
tsplink.trigger[N].stimulus (on page 11-430)

Table of Contents

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
Model3706A
CategorySystem Switch/Multimeter
Voltage Range100 nV to 300 V
Basic DC Voltage Accuracy0.0035%
Basic Resistance Accuracy0.01%
Maximum Channels576
Maximum Voltage300 V
ConnectivityLAN, USB, GPIB
InterfaceLAN, USB, GPIB
Switching MatrixYes
Channels576 (6-slot mainframe)
Maximum Current3 A

Related product manuals