Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-291
localnode.serialno
This attribute stores the instrument's serial number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
serialno = localnode.serialno
serialno
The serial number of the instrument
Details
This indicates the instrument serial number.
Example
display.clear()
display.settext(localnode.serialno)
Clears the instrument display.
Places the serial number of the instrument on the top line
of its display.
Also see
localnode.description (on page 11-284)
localnode.model (on page 11-286)
localnode.revision (on page 11-290)
localnode.showerrors
This attribute sets whether or not the instrument automatically sends generated errors.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Power cycle
Not saved
0 (disabled)
Usage
errorMode = localnode.showerrors
localnode.showerrors = errorMode
errorMode
Show error setting:
â–ª Show errors: 1
â–ª Do not show errors: 0
Details
If this attribute is set to 1, the instrument automatically sends any generated errors stored in the error
queue, and then clears the queue. Errors are processed after executing a command message (just
before issuing a prompt, if prompts are enabled).
If this attribute is set to 0, errors are left in the error queue and must be explicitly read or cleared.
When using this command from a remote node, localnode should be replaced with the node
reference, for example, node[5].showerrors.

Table of Contents

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
Model3706A
CategorySystem Switch/Multimeter
Voltage Range100 nV to 300 V
Basic DC Voltage Accuracy0.0035%
Basic Resistance Accuracy0.01%
Maximum Channels576
Maximum Voltage300 V
ConnectivityLAN, USB, GPIB
InterfaceLAN, USB, GPIB
Switching MatrixYes
Channels576 (6-slot mainframe)
Maximum Current3 A