EasyManuals Logo

Keithley 3706A Reference Manual

Keithley 3706A
936 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #914 background imageLoading...
Page #914 background image
Series 3700A System Switch/Multimeter Reference Manual Appendix C: Status model
3700AS-901-01 Rev. D/June 2018 C-9
Questionable summary bit (Questionable event register)
The summary bit of the questionable event register provides enabled summary information to Bit B3
(QSB) of the status byte.
Figure 183: Questionable event register
As shown above, there is only one register set associated with the questionable status. Attributes are
summarized in status.questionable.* (on page 11-385). Keep in mind that bits can also be set by
using numeric parameter values. For details, see Programming enable and transition registers (on
page C-14).
For example, any of the following statements will set the thermal aspect enable bit of a card in slot 1:
status.questionable.enable = status.questionable.S1THR
status.questionable.enable = status.questionable.SLOT1_THERMAL
status.questionable.enable = 512
The following command will request the questionable enable register value in numeric form:
print(status.questionable.enable)
The bits used in this register set are described as follows:
• SxTHR: Set bit indicates the thermal aspect of the card in slot x is in question, where x = 1 to 6.

Table of Contents

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
Model3706A
CategorySystem Switch/Multimeter
Voltage Range100 nV to 300 V
Basic DC Voltage Accuracy0.0035%
Basic Resistance Accuracy0.01%
Maximum Channels576
Maximum Voltage300 V
ConnectivityLAN, USB, GPIB
InterfaceLAN, USB, GPIB
Switching MatrixYes
Channels576 (6-slot mainframe)
Maximum Current3 A

Related product manuals