Example
format.asciiprecision = 10
x = 3.14159265
format.data = format.ASCII
printnumber(x)
format.data = format.REAL64
printnumber(x)
Output a number represented by x in ASCII
using a precision of 10, then output the same
number in binary using double precision
format.
Output:
3.141592650e+00
#0ñÔÈSû! @
Also see
format.asciiprecision (on page 11-245)
format.byteorder (on page 11-246)
printbuffer() (on page 11-299)
printnumber() (on page 11-302)
fs.chdir()
This function sets the current working directory.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
workingDirectory = fs.chdir("path")
workingDirectory
Returned value containing the working path
path
A string indicating the new working directory path
Details
The new working directory path may be absolute or relative to the current working directory.
An error is logged to the error queue if the given path does not exist.
Example
testPath = fs.chdir("/usb1/")
Change the working directory to usb1.
Also see
None

Table of Contents

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
Model3706A
CategorySystem Switch/Multimeter
Voltage Range100 nV to 300 V
Basic DC Voltage Accuracy0.0035%
Basic Resistance Accuracy0.01%
Maximum Channels576
Maximum Voltage300 V
ConnectivityLAN, USB, GPIB
InterfaceLAN, USB, GPIB
Switching MatrixYes
Channels576 (6-slot mainframe)
Maximum Current3 A

Related product manuals