Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-325
scan.scancount
This attribute sets the number of times the scan is repeated.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Reset
Scan reset
Recall setup
Create configuration script
Save setup
1
Usage
scanCount = scan.scancount
scan.scancount = scanCount
scanCount
The scan count value (1 to 2,000,000,000)
Details
The scan count attribute setting indicates how many times the scan list is iterated through before the
scan completes.
During a scan, the instrument iterates through the arm layer of the trigger model the specified number
of times. After performing the specified number of iterations, the instrument returns to an idle state.
Example
scan.create("1001:1009")
scan.scancount = 100
scan.scaninterval = 1.0 -- delay between scans
Create a scan that includes channels 1 to 9 of slot 1.
Set the scan count to 100.
Set the delay between scans to 1 s.
Also see
Scanning and triggering (on page 5-23)
Trigger model (on page 5-24)

Table of Contents

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
Model3706A
CategorySystem Switch/Multimeter
Voltage Range100 nV to 300 V
Basic DC Voltage Accuracy0.0035%
Basic Resistance Accuracy0.01%
Maximum Channels576
Maximum Voltage300 V
ConnectivityLAN, USB, GPIB
InterfaceLAN, USB, GPIB
Switching MatrixYes
Channels576 (6-slot mainframe)
Maximum Current3 A