Series 3700A System Switch/Multimeter Reference Manual Appendix B: Verification and adjustment
3700AS-901-01 Rev. D/June 2018 B-31
AC adjustment step 12: 1 mA at 1 kHz
1. Send the following command:
dmm.range = 1e-3
2. Source 1 mA at 1 kHz.
3. Send the following command:
dmm.calibration.ac(12) print("done")
AC adjustment step 13: 10 mA at 1 kHz
1. Send the command:
dmm.range = 10e-3
2. Source 10 mA at 1 kHz.
3. Send the command:
dmm.calibration.ac(13) print("done")
AC adjustment step 14: 100 mA at 1 kHz
1. Send the command:
dmm.range = 100e-3
2. Source 100 mA at 1 kHz.
3. Send the command:
dmm.calibration.ac(14) print("done")
AC adjustment step 15: 1 A at 1 kHz
1. Send the command:
dmm.range = 1
2. Source 1 A at 1 kHz.
3. Send the command:
dmm.calibration.ac(15) print("done")
AC adjustment step 16: 2 A at 1 kHz
1. Send the command:
dmm.range = 2
2. Source 2 A at 1 kHz.
3. Send the command:
dmm.calibration.ac(16) print("done")

Table of Contents

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
Model3706A
CategorySystem Switch/Multimeter
Voltage Range100 nV to 300 V
Basic DC Voltage Accuracy0.0035%
Basic Resistance Accuracy0.01%
Maximum Channels576
Maximum Voltage300 V
ConnectivityLAN, USB, GPIB
InterfaceLAN, USB, GPIB
Switching MatrixYes
Channels576 (6-slot mainframe)
Maximum Current3 A

Related product manuals