Series 3700A System Switch/Multimeter Reference Manual Appendix B: Verification and adjustment
3700AS-901-01 Rev. D/June 2018 B-33
Save adjustments
Program today's date, calibration due date, and serial number, and save the calibration constants in
EEPROM (electrically erasable programmable read-only memory) by sending the following
commands:
dmm.adjustment.date=os.time()
dmm.calibration.save()
dmm.calibration.verifydate=dmm.adjustment.date
dmm.calibration.lock()
dmm.reset()
Adjustments are complete after they have been saved and locked.

Table of Contents

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
Model3706A
CategorySystem Switch/Multimeter
Voltage Range100 nV to 300 V
Basic DC Voltage Accuracy0.0035%
Basic Resistance Accuracy0.01%
Maximum Channels576
Maximum Voltage300 V
ConnectivityLAN, USB, GPIB
InterfaceLAN, USB, GPIB
Switching MatrixYes
Channels576 (6-slot mainframe)
Maximum Current3 A