EasyManua.ls Logo

Keithley 6487 - Ground Loops and Measurement Errors

Keithley 6487
338 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 6487 Reference Manual Buffer and Sweeps 6-7
F) :FORMat:ELEMents <list>
List parameters:
READing — Includes the buffer reading in each data string.
UNITs — Identifies amps, ohms, or math units.
VSOurce — Includes voltage source value in each reading string.
TIME — Includes the timestamp for each reading. Timestamp can be in the abso-
lute or delta format (see Ref D).
STATus — Includes a status word for each reading. It provides status information
on instrument operation. (See “FORMat subsystem,” page 13-4.)
DEFault — All except VSource.
ALL — Includes all reading elements.
At least one data element must be in the list. Listed elements must be separated by
a comma (i.e. FORMat:ELEMents, READing, TIME). Elements not listed will not
accompany the response message for TRACe:DATA?. Data in the response mes-
sage will be in order of the listed data elements set by this command.
G) :CALCulate3:FORMat <name>
This command selects the statistic to be returned by CALCulate3:DATA? (see Ref H).
Name parameters:
MINimum — Select the lowest reading stored in the buffer.
MAXimum — Select the largest reading stored in the buffer.
MEAN — Select the mean average statistic for the readings stored in the buffer.
SDEViation — Select the standard deviation statistic for the readings stored in the
buffer.
PKPK — Select the peak-to-peak statistic for readings stored in the buffer.
Peak-to-Peak is calculated as follows: PKPK = MAXimum - MINimum.
H) :CALCulate3:DATA?
1. If the number of data points in the buffer is one or none, CALCulate3:DATA? will
result in an error (-230).
2. If there is a lot of data in the buffer, some statistic operations may take too long and
cause a bus time-out error. To avoid this, send CALC3:DATA? and then wait for
the MAV (message available) bit in the Status Byte Register to set before address-
ing the Model 6487 talk (Section 10).

Table of Contents

Other manuals for Keithley 6487

Related product manuals