10-20 Status Structure Model 6487 Reference Manual
Programming example — read error queue
The following command reads the error queue:
STAT:QUE? ' Read Error Queue.
Table 10-7
SCPI commands — error queue
Command Description Default
STATus
:QUEue
[:NEXT]?
:ENABle <list>
:ENABle?
:DISable <list>
:DISable?
:CLEar
SYSTem
:ERRor
[:NEXT]?
:ALL?
:COUNt?
:CODE
[:NEXT]?
:ALL?
STATus subsystem:
Read error queue:
Read and clear oldest error/status (code and message).
Specify error and status messages for error queue.
Read the enabled messages.
Specify messages not to be placed in queue.
Read the disabled messages.
Clear messages from error queue.
SYSTem subsystem:
Read error queue:
Read and clear oldest error/status (code and message).
Read and clear all errors/status (code and message).
Read the number of messages in queue.
Code numbers only:
Read and clear oldest error/status (code only).
Read and clear all errors/status (codes only).
(Note 1)
(Note 2)
(Note 2)
(Note 1)
Notes:
1. Power-up and *CLS empties the error queue. STATus:PRESet has no effect.
2. Power-up enables error messages and disables status messages. *CLS and STATus:PRESet have no effect.

Table of Contents

Other manuals for Keithley 6487

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6487 and is the answer not in the manual?

Keithley 6487 Specifications

General IconGeneral
TypePicoammeter/Voltage Source
Current Measurement Range20 fA to 20 mA
InterfacesGPIB, RS-232
Voltage Burden< 200 µV
Input Impedance>10^14 Ω
Power100 V / 120 V / 220 V / 240 V
Basic Current Measurement Accuracy0.4%

Related product manuals