12-4 SCPI Signal Oriented Measurement Commands Model 6487 Reference Manual
D) MEASure[:<function>]? Configure and perform “one-shot” measurement
<function> = CURRent[:DC] Measure current
This command combines all of the other signal oriented measurement commands to per-
form a “one-shot” measurement and acquire the reading.
When this command is sent, the following commands execute in the order that they are
presented.
CONFigure:<function>
READ?
When :CONFigure is executed, the instrument goes into a “one-shot” measurement mode.
See CONFigure for details.
When READ? is executed, its operations will then be performed. In general, an INITiate is
executed to perform the measurement and a FETCh? is executed to acquire the reading.
See :READ? for details.

Table of Contents

Other manuals for Keithley 6487

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6487 and is the answer not in the manual?

Keithley 6487 Specifications

General IconGeneral
TypePicoammeter/Voltage Source
Current Measurement Range20 fA to 20 mA
InterfacesGPIB, RS-232
Voltage Burden< 200 µV
Input Impedance>10^14 Ω
Power100 V / 120 V / 220 V / 240 V
Basic Current Measurement Accuracy0.4%