MRS-BA-defsi-0410 104
Procedure for changing the password:
• Enter the password and press
«
».
DATA-PROTECTION MED
•
The current data protection
status is now displayed.
•
Press «
».
new PASSWORD 6 1 0 0
• Enter the new password,
• (e.g. 6123, procedure as
described above).
new PASSWORD
6
1
2 3
• Press «
» to save the entry.
4.4.10 Anti-theft encoding
The instrument can be protected against theft by using a freely
selectable, four-digit numerical code:
•
If the anti-theft code is deactivated, the instrument can be
switched on again and operated after a power cutoff without
having to enter a code.
•
If the anti-theft code is activated, the instrument asks for the
code to be input after each power cutoff.
•
If the code is entered incorrectly, the instrument is blocked.
•
If the instrument is blocked, it must first be disconnected
from the power supply, then reconnected and unblocked by
entering the correct code.
•
After seven consecutive incorrect entries, the display reads
“NO ACCESS, CALL SERVICE”. In this case only a KERN
service engineer can unblock the instrument again.
NOTE
The anti-theft encoding is deactivated in the factory settings.
The
preprogrammed code
set in the factory is:
8 9 3 7
This code is the same in all KERN instruments. Therefore, for
security reasons enter your own code.
Keep your