7.3 Calibration
The calibration of the MRS 120-3 moisture analyzer is defined in
the configuration menu (see chapter 3.8"Weight calibration" and
chapter 4.4.5 “Balance calibration”).
NOTE
The balance calibration and the temperature calibration can be
interrupted at any time by pressing «ON/OFF».
7.3.1 Calibrating the balance
Possible types of balance calibration:
External calibration by means of ICM (Intelligent Calibration
Mode)
External calibration with a freely definable weight
External calibration by means of ICM
Calibration weights in increments of 10g can be used for the MRS
120-3 moisture analyzer whereby the calibration weights must
correspond to the precision of the instrument.
"SET CALIBRATION MODE EXTERNAL" must be set in the
configuration menu (see chapter 4.4.5"Balance calibration") for
external calibration by means of ICM.
Ensure that there is no drying
taking place, i.e. that the
instrument is in weighing mode.
Balance calibration
Keep pressing «TARE» until
"BALANCE CALIBRATION" is
displayed.
- - 0000 9
The instrument performs a zero
point measure (0000 g is
displayed flashing).
- - 50 9
After the zero point
measurement, the display
flashes the recommended
calibration weight.
Put the calibration weight on.