121
MRS-BA-defsi-0410
5.3 Setting drying parameters
The drying parameter setting procedure is started with the four
function keys under the display.
Each of the four function keys starts the entry of the corresponding
drying parameter. The parameters are entered or changed in the
same way as the menu operation procedure ( except that only the
current parameter can be changed each time.
5.3.1 Heating program
This function key is used to start the heating program selection.
There are three heating programs available for determining the
moisture content:
Standard drying
Boost drying (0.1Min – 99.9Min)
Soft drying
HEATING PROGRAM
HEAT MODE STANDARD
BOOST
SOFT
Select the heating program
BOOST-time selectable*
* The BOOST-Time can be set in 0.1 min increments from 0.1 min
... 99.9 min.
Standard drying
The drying temperature is predefined by the user. The end
temperature is started up with a high heat output and is kept
constant with slight overshooting.
This program is used for most samples.