This program is used for samples with a low moisture content in
which there is a risk of combustion.
5.3.2 Temperature
This function key is used to start the drying temperature input.
The drying temperature can be entered in 1°C increments from
30°C ... 230°C.
After 10 minutes temperatures higher than 200°C are automatically
adjusted down to 200°C over the course of the next 20 minutes.
DRYING TEMPERATURE
TEMPERATURE
105
°C Temperature input
The temperature setting for drying with the MRS 120-3 is lower
than in the case of drying with the drying oven technique.
5.3.3 Timer stop
This function key is used to define the drying time.
If the timer stop is switched on, the measurement procedure is
ended once the time set has elapsed.
The time can be set in 0.1 min increments from 0.1 min ... 240.0
min.
DRYING TIME
STOP TIME
10.0
MIN Only if the timer stop is switched on
TIMER STOP ON/
OFF