125
MRS-BA-defsi-0410
One digit is the smallest change in measurement that can be
displayed by the moisture analyzer.
In the MRS 120-3: 1 digit = 1 mg
ADAPTSTOP
Is a fully automatic stop mode which determines the switch-off time
on the basis of the drying progress.
5.4 Application table (recommendations)
Preparation of the standard sample:
-
Distribute the sample evenly on the aluminium tray.
-
Reduce the sample to small pieces and distribute evenly
on the aluminium tray.
Preparation of special samples:
-
A glass fibre filter can be used if the test material is
sensitive
or difficult to distribute (e.g. mercury).
-
Apply the sample evenly to the glass fibre filter and cover
with a second glass fibre filter.
-
The glass fibre filter can also be used for protection against
material splashes (every splash adulterates the final
result).