Liquids:
You are advised to use a fiber glass filter for liquids, pastes or
slurry samples. The fiber glass filter offers the following
advantages:
even distribution on account of the capillary effect
no formation of drops
quick evaporation due to the larger surface
5.2.1 Preventing samples being encrusted
In order to avoid the sample becoming encrusted, solvent can also
be added to the sample after the measurement has started. The
solvent added has no bearing on the final result of the
measurement.
Start the measurement, automatically or by pressing the
«
START/STOP
» key.
The dryer hood can be opened again within 5 seconds of the
start. During this time, the words "START DRYING" are
displayed in the info line of the display.
After opening the sample chamber, you can add additional
solvent at any time until the hood is closed. Once the dryer
hood is closed, the measurement is continued. "START
COVER CLOSE" appears in the info line in the display. If you
press the «START/STOP» key, the measurement is
interrupted.
NOTE
The additional solvent is taken into account in the measurement
printout because all the interim values are calculated on the basis of
the current weight value.
However, it has no bearing on the drying result because the solvent
has completely dried off.