4.6 Special operating keys
4.6.1 The tare key
–
Ensure that there is no drying taking place, i.e. that the
instrument is in weighing mode.
• Activating taring
–
Press «
TARE
» briefly.
– A taring procedure is performed.
• Activating calibration
– Keep «TARE» pressed until "BALANCE CALIBRATION" is
displayed.
– Release «TARE».
–
The balance carries out a calibration in accordance with the
settings in the configuration menu and logs these in the form of
a printout (see chapter 4.4.5"Balance calibration" and chapter
7.3.1 “Calibrating the balance”)
• Activating a calibration test
– Keep «TARE» pressed until "TEST CALIBRATION" is
displayed.
– Release «TARE».
–
A calibration test is performed on the balance.
• Activating a temperature calibration
– Keep «TARE» pressed until "TEMP. CALIBRATION" is
displayed.
– Release «TARE».
–
The MRS 120-3 moisture analyzer carries out a temperature
calibration and logs it in the form of a printout (see chapter
4.4.6"Balance calibration" and chapter 7.3.2 “Calibrating the
balance”).
• Activating a temperature calibration test
– Keep «
TARE
» pressed until "TEST TEMPERATURE" is
displayed.
– Release «
TARE
».
– The MRS 120-3 moisture analyzer carries out a temperature
test.