MRS-BA-defsi-0410 130
Command Function
SDTmmddyy
hhmmss
Set date and time (Month, Day, Year, Hour, Minutes,
Seconds)
T (ttt) Tare or set tare to a specific value
ZERO
Set instrument to 0 (if weight is stable and within the
zero setting range)
Rttt
Adjusts the heating to the required temperature
(30°C ... 230°C)
ROFF Switch off heating
PWT (ttt.t)
Print weight value and temperature value
Print with time basis in s (ttt.t)
(switch off by transmitting PWT)
NOTE
Each remote control command must terminate with «CR» «LF».
The commands are acknowledged if required.
6.2.1 Examples of the remote control of the instr.
Input Description of the function activated
D – – – – – Five dashes are displayed
DTEST123 tESt123 is displayed
D The display is dark
T10 -10.000 g (Tare set to = 10 g)
T1 -1.000 g (Tare set to = 1g)
T Instrument is tared
R100 Adjusts the temperature to 100 °C