4.5.5 Print rate
• PRINT RATE
PRINT RATE
1.0
MIN
Print rate interval
The interval at which interim results are printed is set at "PRINT
RATE". The print interval can be set in 0.1 min increments from
0.1 ... 10.0 min.
4.5.6 Standby temperature
• STANDBY TEMP.
STANDBY TEMP. ON/
OFF
Switch standby temperature on/off
TEMPERATURE 40°C
Temperature value, only if the standby
temperature is switched on
Adjusts the temperature in the sample chamber to the set
temperature, provided the sample chamber is closed. The
temperature range available is 30°C ... 100°C. A flashing circle in
the display indicates the stanby temperature is not reached, yet.
4.5.7 Autostart
• AUTOSTART
AUTOSTART ON/OFF
If autostart is switched on, the measurement is started as soon as
the sample chamber is closed. This is provided that the moisture
analyzer is prepared for a new measurement.