MRS-BA-defsi-0410 124
5.3.4 Switch-off criteria
This function key is used to start the definition of the switch-off
criteria for the measurement.
One firmly and 2 freely definable stop mode are availlable.
There is also the automatic stop mode "ADAPTSTOP".
– Digit / Time (selectable)
–
Prozent / Time (selectable)
– ADAPTSTOP
• STOPMODE
AUTOSTOP
OFF
Digit / Time
Percent / Time
ADAPTSTOP
Switch-off criterion Digit per time
The drying procedure is switched off as soon as the reduction in
weight, during the time set, is smaller than the number of digits set.
The reduction in weight must have been greater than the switch-off
criterion at some point.
In the freely definable mode, the user can set from 1 ... 99 digits in
1 digit increments and from 10...90 seconds in 10 second
increments.
Switch-off criterion percent per time
The drying procedure is switched off as soon as the reduction in
weight, during the percent set, is smaller than the number of digits
set. The reduction in weight must have been greater than the
switch-off criterion at some point.
In the freely definable mode, the user can set from 1 ... 99 digits in
1 digit increments and from 10...90 seconds in 10 second
increments.