MRS-BA-defsi-0410 122
Boost drying
The drying temperature is predefined by the user. During the
selected time (0.1Min-99.9Min) of drying, the target temperature is
exceeded by 40%. Once this time has elapsed, the temperature is
adjusted down to the target temperature. The temperature is
started up with a high heat output.
The maximum temperature reached during the boost is 230°C.
This program is used for samples with a very high moisture
content.
Soft drying
The drying temperature is predefined by the user. The end
temperature is started up gently with a low heat output. The end
temperature is reached after approx. 4 minutes.