MRS-BA-defsi-0410 134
- - 50 9
The display flashes.
+ 50 9
Once the display stops
flashing,this indicates that the
calibration has ended.
External calibration with a freely definable weight
"SET CALIBRATION MODE EXT.-DEF." must be set in the
configuration menu (se chapter 4.4.5 “Balance calibration”) for
external calibration with a freely definable weight.
You then need to enter the effective value of the calibration weight
(DEF. n,nnnn g) with up to ten times the precision of the instrument
balance.
NOTE
Is calibrated with the free weight, so only this weight may be used.
Procedure to follow:
Ensure that there is no drying
taking place, i.e. that the
instrument is in weighing mode.
Balance calibration
Keep pressing «
TARE
» until
"BALANCE CALIBRATION" is
displayed.
- - 0000 9
The instrument performs a zero
point measure (0000 g is
displayed flashing).
- - 22 9
After the zero point
measurement the display
flashes the calibration weight
entered beforehand.
Put the calibration weight on.
- - 22 9
The display flashes.
+ 22.150 9
Once the display stops
flashing, this indicates that the
calibration has ended (the
precise value is displayed).