EasyManua.ls Logo

KERN MRS120-3

KERN MRS120-3
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MRS-BA-defsi-0410 136
The temperature value is
transmitted to the instrument
from the temperature
adjustment set.
TEMP 100°C 20.00 MIN
It is heated to 160°C.
The temperature value is
transmitted to the instrument
from the temperature
adjustment set.
Once the temperature
calibration has ended, the
report is printed.
Temperature calibration report printout
Temperature Calibration
-----------------------
Temperature calibration report
Date 16.10.2003 Time 12:51:36
Name : MRS 120-3
Software : N50-0000 P00
Serialno : 3300-1
Time of the calibration and instrument data
Temp. Reference ID : ...
Identifier of the temperature adjustment set
Temperature 100 C : 100 C
Temperature 160 C : 160 C
Status of the temperature calibration
Temperature Calibration o.k.
Instrument has been correctly temperature-
calibrated
Operator :
Operator ID, if activated under Set Print
Format (see chapter 4.4.3 Configuring the
report printout).

Table of Contents

Related product manuals