MRS-BA-defsi-0410 138
Fault
Possible causes
Results of
weighing are
clearly incorrect
The instrument has not been correctly tared
The instrument has not been correctly levelled
The calibration is no longer correct
Sharp temperature fluctuations occur
The weight
display changes
continuously
The draft is too strong at the location of the
instrument
The instrument support is vibrating or
fluctuating
The weighing pan is touching a foreign body
The sample is absorbing moisture
The sample is
evaporating/vaporising/subliming
Sharp changes in temperature in the sample
Configuration
menu cannot be
changed
The password lock is activated in the
configuration menu
The display
flashes
continuously
during
calibration
The instrument location is not quiet enough
(interrupt calibration with «ON/OFF» and
move the balance to a better location).
Use of a calibration weight which is too
imprecise (only applies to external calibration)
The connected
printer won't
work
The printer is not switched on
The data cable is faulty or not connected
The interface settings do not match the
moisture analyzer
The printer
prints incorrect
characters
The parity setting or the baud rate of the
interface don't match
The data cable is faulty
Drying won't
start
The sample is not stable