Fault
Possible causes
Results of
weighing are
clearly incorrect
•
The instrument has not been correctly tared
•
The instrument has not been correctly levelled
•
The calibration is no longer correct
•
Sharp temperature fluctuations occur
The weight
display changes
continuously
•
The draft is too strong at the location of the
instrument
• The instrument support is vibrating or
fluctuating
•
The weighing pan is touching a foreign body
•
The sample is absorbing moisture
•
The sample is
evaporating/vaporising/subliming
•
Sharp changes in temperature in the sample
Configuration
menu cannot be
changed
• The password lock is activated in the
configuration menu
The display
flashes
continuously
during
calibration
•
The instrument location is not quiet enough
(interrupt calibration with «ON/OFF» and
move the balance to a better location).
• Use of a calibration weight which is too
imprecise (only applies to external calibration)
The connected
printer won't
work
•
The printer is not switched on
•
The data cable is faulty or not connected
• The interface settings do not match the
moisture analyzer
The printer
prints incorrect
characters
•
The parity setting or the baud rate of the
interface don't match
•
The data cable is faulty
Drying won't
start
•
The sample is not stable